MS-250 Mixed Signal Oscilloscope Option
Qualified Trigger
Qualified arms the trigger on the A event, then fires on the B event. In Normal trigger mode,
it automatically resets after the B event. The options for the B event depend on the type of
A event. You may apply additional Holdoff by time or number of events.
On the Trigger dialog, select Qualified or MultiStage > Qualified trigger type to display the
controls, depending on your oscilloscope.
Then, on the Qualified dialog choose the A and B events.
Besides an Edge or Pattern trigger, two special conditions may be selected as the arming
("A") event:
l State, voltage measured above or below a threshold Level.
l PatState, a pattern that persists over a user-defined number of events or time. Like
Pattern triggers, PatState events may be analog voltage patterns, digital logic
patterns, or a mix of both, depending on the oscilloscope's capabilities.
NOTE: On a standard oscilloscope, Pattern and PatState events will default to the analog
pattern setup dialog. On a Mixed-Signal oscilloscope, Pattern and PatState events will
default to the digital pattern setup dialog.
Once you've selected the A and B events on the Qualified dialog, set up the conditions on
the respective sub-dialogs exactly as you would a single-stage trigger.