M
81331 Rev 01 Nov 2010
16
HV ON Warning signal indicating that the HV unit is active
OPEN
GROUND
When lit, this yellow lamp indicates an open in
double ground system or defective grounding of test
set.
SAFETY
INTERLOCK
1 and 2
Two plug receptacles for connecting external
interlock switches. Two hand interlock switches and a
foot switch are supplied; however, in the event that a
hand interlock is replaced with a test area interlock,
the system must be constructed so that the interlock
switches are closed when the test area gate or gates
are closed. The interlock wiring must be run as a
twisted pair to minimize electromagnetic coupling
into the system. This interlock system should be
wired such that connection is made to the A and B
sockets of the SAFETY INTERLOCK receptacle.
When the interlock loop is opened the test is
automatically terminated.
INPUT RED Plug receptacle for connecting the red low-voltage
test lead.
INPUT
BLUE
Plug receptacle for connecting the blue low-voltage
test lead.
USB port Receptacle for computer communication.
Ethernet port Receptacle for computer communication.
Ext-Int PC
switch
Switch for selecting internal or external PC
(functional on Delta4310 only, disabled on
Delta4110). “Internal” connects the on-board PC to
Delta via Ethernet, “External” means that you can
operate Delta from an external computer.