Delta 4000 Manuel d'instructions Page 44

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 108
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 43
M
81331 Rev 01 Nov 2010
36
Excitation Current
Here you can select either frequency variation or line sync reversal (default) for
noise suppression
Auto tip-up Test
Set the kV/step for the auto tipup test. The software will automatically increase
the voltage by the increment set in this box until the entered test voltage selected
in the main screen is reached.
You can also select either frequency variation or line sync reversal for noise
suppression
Results
Specify if you want the measured values presented as either 10 kV equivalent or
the actual measured values.
Frequency Sweep
Set frequencies at which the frequency sweep will be made.
Manual
Select the rejection mode to be used during manual measurements.
Language
Set the language for the user interface of the software.
Vue de la page 43
1 2 ... 39 40 41 42 43 44 45 46 47 48 49 ... 107 108